• ASTM D5542-04(2009)

ASTM D5542-04(2009)

  • standard by ASTM International, 05/01/2009
  • Standard Test Methods for Trace Anions in High Purity Water by Ion Chromatography
  • Category: ASTM

$60.00 $30.00

Full Description

1.1 These test methods cover the determination of trace (μg/L) levels of fluoride, acetate, formate, chloride, phosphate, and sulfate in high purity water using ion chromatography in combination with sample preconcentration. Other anions, such as bromide, nitrite, nitrate, sulfite, and iodide can be determined by this method. However, since they are rarely present in significant concentrations in high purity water, they are not included in this test method. Two test methods are presented and their ranges of application, as determined by a collaborative study, are as follows:

Range Tested
(μg/L Added)
Limit of DetectionA
(Single Operator)
(μg/L)
Sections
Test Method A: 7-15
Chloride 0-24 0.8
Phosphate 0-39 B
Sulfate 0-55 1.8
Test Method B: 16-23
Fluoride 0-14 0.7
Acetate 0-414 6.8
Formate 0-346 5.6

A Limit of detection is lowest measurable concentration not reportable as zero at 99 % level of confidence as per EPRI study as cited in Sections 15 and 23.

B Insufficient data to calculate limit of detection.

1.2 It is the user's responsibility to ensure the validity of these test methods for waters of untested matrices.

1.3 The common practical range of Test Method A is as follows: chloride, 1 to 100 μg/L, phosphate, 3 to 100 μg/L, and sulfate, 2 to 100 μg/L.

1.4 The common practical range of Test Method B is as follows: fluoride, 1 to 100 μg/L, acetate, 10 to 200 μg/L, and formate, 5 to 200 μg/L.

1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

7.1 This test method is optimized for the quantitative determination of trace levels of chloride, phosphate, and sulfate. Anions such as fluoride, acetate, and formate can be detected by this method, but are not reliably resolved from each other. See Fig. 1 for a typical chromatogram.

7.2 Using a concentrated sample volume of 20 mL, the test method is applicable in the range outlined in Section 1. The range of this test method may be extended by concentrating a smaller or a larger sample volume. Be sure not to exceed concentrator column breakthrough volume (see annex).

16.1 This test method is optimized for the quantitative determination of trace levels of fluoride, acetate, and formate in high purity water and in water containing up to 2000 mg/L boron from bo

Product Details

Published:
05/01/2009
Number of Pages:
9
File Size:
1 file , 150 KB
Redline File Size:
2 files , 270 KB
Note:
This product is unavailable in Russia, Ukraine, Belarus

Document History

ASTM D5542-04(2009)

ASTM D5542-04

ASTM D5542-94(1999)e1

ASTM A897M-90

ASTM A897M-90

Standard Specification for Austempered Ductile Iron Castings [Metric]..

$17.00 $33.00

ASTM D3540-90(1998)

ASTM D3540-90(1998)

Standard Specification for Primary Amyl Acetate, Synthetic (98 % Grade)..

$27.00 $54.00

ASTM D3533-90(2001)

ASTM D3533-90(2001)

Standard Test Method for Rubber - ;Nitrogen Content..

$26.00 $52.00

ASTM D3533-90(1995)e1

ASTM D3533-90(1995)e1

Standard Method of Rubber - ;Nitrogen Content..

$27.00 $54.00