• ASTM D7315-17

ASTM D7315-17

  • standard by ASTM International, 07/01/2017
  • Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode
  • Category: ASTM

$75.00 $38.00

Full Description

1.1 This test method covers the static determination of turbidity in water. Static refers to a sample that is removed from its source and tested in an isolated instrument. (See Section 4.)

1.2 This test method is applicable to the measurement of turbidities greater than 1.0 turbidity unit (TU). The upper end of the measurement range was left undefined because different technologies described in this test method can cover very different ranges. The round robin study covered the range of 0-4000 turbidity units because instrument verification in this range can typically be covered by standards that can be consistently reproduced.

1.3 Many of the turbidity units and instrument designs covered in this test method are numerically equivalent in calibration when a common calibration standard is applied across those designs listed in Table 1. Measurement of a common calibration standard of a defined value will also produce equivalent results across these technologies.

Design and
Reporting Unit

Prominent Application

Key Design Features

Typical
Instrument Range

Suggested
Application Ranges

Nephelometric non-ratio (NTU)

White light turbidimeters. Comply with U.S. EPA Method 180.1 for low level turbidity monitoring.

Detector centered at 90° relative to the incident light beam. Uses a white light spectral source.

0.0-40

0.0-40 Regulatory

Ratio White Light turbidimeters (NTRU)

Complies with ISWTR regulations and Standard Method 2130B. Can be used for both low and high level measurement.

Used a white light spectral source. Primary detector centered at 90°. Other detectors located at other angles. An instrument algorithm uses a combination of detector readings to generate the turbidity reading.

0-10 000

0-40 Regulatory
0-10 000 other

Nephelometric, near-IR turbidimeters, non-ratiometric (FNU)

Complies with ISO 7027. The wavelength is less susceptible to color interferences. Applicable for samples with color and good for low level monitoring.

Detector centered at 90° relative to the incident light beam. Uses a near-IR (780-900 nm) monochromatic light source.

0-1000

0-40 Regulatory (non-US)
0-1000 other

Nephelometric near-IR turbidimeters, ratio metric (FNRU)

Complies with ISO 7027. Applicable for samples with high levels of color and for monitoring to high turbidity levels.

Uses a near-IR monochromatic light source (780-900 nm). Primary detector centered at 90°. Other detectors located at other angles. An instrument algorithm uses a combination of detector rea

Product Details

Published:
07/01/2017
Number of Pages:
22
File Size:
1 file , 700 KB
Redline File Size:
2 files , 1.8 MB
Note:
This product is unavailable in Russia, Ukraine, Belarus

Document History

ASTM D7315-17

ASTM D7315-07a

ASTM D7315-07

ASTM C412-11

ASTM C412-11

Standard Specification for Concrete Drain Tile..

$30.00 $60.00

ASTM D3262-11

ASTM D3262-11

Standard Specification for "Fiberglass" (Glass-Fiber-Reinforced Thermosetting-Resin) Sewer Pipe..

$30.00 $60.00

ASTM D3364-99(2011)

ASTM D3364-99(2011)

Standard Test Method for Flow Rates for Poly(Vinyl Chloride) with Molecular Structural Implications..

$30.00 $60.00

ASTM D516-11

ASTM D516-11

Standard Test Method for Sulfate Ion in Water..

$26.00 $52.00